Configuration

Limits

Note
This tab is only shown for pH and ISE electrodes.

Slope monitoring

on, [ off ]

If this option is switched on then a check will be made whether the

Slope

determined during calibration is within the defined limits. If it is outside them
a corresponding message appears and you can choose whether you will
nevertheless accept the value or whether the previous value is to be re-
tained.

Lower limit

-2.147E+9 ... 2.147E+9

pH:

96.0 %

ISE (+1):

55.0 mV

ISE (-1):

-61.0 mV

ISE (+2):

25.0 mV

ISE (-2):

-31.0 mV

ISE (+3):

16.0 mV

ISE (-3):

-23.0 mV

ISE (+4):

12.0 mV

ISE (-4):

-17.0 mV

Lower limits for the slope.

Upper limit

-2.147E+9 ... 2.147E+9

pH:

101.0 %

ISE (+1):

61.0 mV

ISE (-1):

-55.0 mV

ISE (+2):

31.0 mV

ISE (-2):

-25.0 mV

ISE (+3):

23.0 mV

ISE (-3):

-16.0 mV

ISE (+4):

17.0 mV

ISE (-4):

-12.0 mV

Upper limits for the slope.

pH(0) monitoring

on, [ off ]

If this option is switched on then a check will be made whether the elec-
trode zero point

pH(0)

determined during the calibration of a pH electrode

is within the defined limits. If it is outside them a corresponding message