Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set

threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 1.0 ] ... 20.0 pH/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Additional measured values

Abbildung und Parameter: siehe DET pH - Weitere Messwerte

STAT U

Command for STAT titrations with measured value U.

Devices

This command can be executed with the following devices which have the

STAT

pH

mode:

Titrino: 718, 736, 751, 799

Titrando: 835, 836, 842, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STAT U

are configured in the following 8 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

Start conditions

Parameters for the measurement of the initial measured value and for defin-
ing the conditions which have to be met at the start of the titration.

Control parameters

Parameters for the control point.

Titration parameters

Parameters for the run of the titration.

Stop conditions

Definitions of conditions which cause the stop of the titration.

Monitoring

Definition of measured value, dosing rate, and temperature monitoring.

Advertising