Commands

429

Pause 1

Pause 1

[ 0 ] ... 999999 s

(only Titrando)

Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.

Start volume

Start volume

[ 0.00000 ] ... 9999.99 mL

(Titrando)

[ 0.00 ] ... 999.99 mL

(Titrino)

Volume to be added before the start of the titration with the below entered
speed.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the start volume is to be added. The maximum dosing rate

depends on the cylinder volume of the used exchange unit or dosing unit.

Pause 2

Pause 2

[ 0 ] ... 999999 s

Waiting time, e.g. until the electrode has settled down after the start or a re-
action time after the addition of a start volume. The pause takes place at
the end of all start conditions.

Control parameters

Command name

25 characters, [ STAT pH # ]

Name of the command.

Control point

Control point at pH

-20.000 ... 20.000, [ off ]

(Titrando),

-20.00 ... 20.00, [ off ]

(Titrino)

Definition of the pH value to be controlled.