Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 415

Advertising
background image

Commands

389

Conditioning

Command name

25 characters, [ SET pH # ]

Name of the command.

Conditioning

[ on ], off

Switching conditioning on/off. If conditioning is deactivated all following pa-
rameters are not visible.

Start drift

1 ... [ 20 ] ... 999 µL/min

(only Titrando)

The determination can only be started when the measured drift is below the
entered value. The Status in the Run window (Single determination or De-
termination series) then switches to

COND READY

.

Drift correction

automatic, manual, [ off ]

If this parameter is set to

automatic

or

manual

the drift correction ( =

Drift

value

x Titration time) caused by blank consumption during titration is de-

ducted from the total consumption. With

automatic

the drift value at the start

of the titration applies with

manual

the value defined under

Drift value

is ap-

plied

Drift value

[ 0 ] ... 99.9 µL/min

Drift value for drift correction. This parameter is only shown for

Drift correc-

tion

=

manual

.

Stop volume

0.000000 ... [ 20.00 ] ... 9999.99 mL, [ off ]

(only Titrando)

Stops when the given volume has been added after the start of the titration.
This value should be adapted the volume of the sample used to prevent

overflow.

Stop time

0 ... 999999 s, [ off ]

(only Titrando)

Stops when the given time has elapsed after the start of the titration.

Delay after 'Cond ok'

[ 0 ] ... 999 s

Waiting period until the start of the titration during which the conditioning
condition must be fulfilled.

Advertising