Configuration

Settings according to 21 CFR 11

Sets the parameters to standard settings for FDA-conform work.

Inactivity delay

on, [ off ];

FDA-default:

on

1 ... [ 10 ] ... 60 min;

FDA-default:

10 min

If this option is switched on then the dialog window for signing will be
automatically closed when the entered time limit has expired.

Remove password after signature

on, [ off ];

FDA-default:

on

If this option is switched on then the password must be entered again af-
ter each signature.

Sign each determination separately

on, [ off ];

FDA-default:

on

If this option is switched on then each determination contained in the de-

termination overview must be signed individually.

Default reasons

On the tab

Default reasons

the reasons are defined that have to be entered when

methods and determinations have to be signed for, or when methods, determina-

tions and sample data are altered.

Category

[ Signature level 1 ], Signature level 2, Modification of methods, Modification of
determinations, Modification of sample data

Selection of the category for which the reasons are to be defined.

Reasons

50 characters

Shows the reasons defined for the selected category.

Move text upward (alters sequence).