Method

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Dosing device

Dosing device

[ 1 ] ... 4

(808, 809, 835, 836, 857)

[ 1 ] ... 3

(855)

[ internal D0 ]

(Titrino)

external D1, external D2

(only 736, 751, 758, 799)

Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing.

Solution

24 characters, solution names, [ not defined ]

Input or selection of the titrant or solution defined in the solution table. Addi-
tionally,

not defined

can be selected. If intelligent exchange or dosing units

are used then a check will be made in the method sequence that the cor-
rect titrant is present in the connected dosing device and whether the type
of dosing device is correct. With non-intelligent exchange or dosing units
the cylinder volume is checked. At the start of the command the validity of
the titer and the expiry date of the selected titrant will be checked as well as

the GLP test interval for the buret. If

not defined

is selected no tests will be

carried out.

Sensor

Measuring input

[ 1 ] ... 2

(only 808, 809, 835, 836, 857)

[ 1 ]

(only 855)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ Metal electrode ]

Selection of a sensor of type

Metal electrode

from the list of sensors defined

in the sensor table.

I(pol)

-125.0 ... [ 5.0 ] ... 125.0 µA

(Titrando, in 2.5

µ

A steps)

-127 ... [ 5 ] ... 127 µA

(Titrino, in 1

µ

A steps)

The polarization current is the current applied to the polarized electrode
during a voltametric measurement.

Electrode test

on, [ off ]

Switching on/off the electrode test for polarized electrodes. The electrode

test for polarized electrodes is carried out during the transition from an in-
active condition to a measuring condition.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured

continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.