Method

Measuring parameters

The measuring parameters are defined on this tab.

Command name

25 characters, [ MEAS pH # ]

Name of the command.

Measurement

The two measuring modes Measurement with drift control or Meas-
urement without drift control
are selectable.

Note
A constant measured value is often only achieved after a certain time, as mixing
and possibly the reaction itself require a certain time. The response time of an
electrode can also increase with time, i.e. reaching a constant measured value

takes longer and longer. In such a case

drift-controlled measurement is par-

ticularly advisable, as the measured values are only accepted when equilibrium
has almost been achieved.

Measurement with drift control

[ on ], off

The measurement is carried out drift-controlled if this option is enabled. The
measurement is stopped as soon as the

Signal drift

or the

Stop measured

value pH

is reached or a

waiting time

has elapsed. The corresponding pa-

rameters are not visible if this option is disabled.

Signal drift

0.1 ... [ 10.0 ] ... 999.0 mV/min

(Titrando)

0.5 ... [ 10.0 ] ... 999.0 mV/min

(Titrino)

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Min. waiting time

[ 0 ] ... 999'999 s

(Titrando only)

The measured value is only accepted if the minimum waiting time has
elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.