72

General program functions

If the output is to be an impulse with a different length then an appropriate
template must be defined for it.
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Examples:

************1*

sets output line 1 to active (= set) which would corre-

spond to a Stop command, e.g. for a connected Titrino.

************0*

sets the line to inactive.

Note

We recommend that non-relevant output lines are masked with an asterisk

*

so

that the status of these lines is not altered.

Note

For Titrinos with 3 output lines only 3 characters can be entered. If a signal
template is selected, only the first 3 characters are used.

Set the binary pattern defined under

Outputs

.








2.6.8

Sample changer functions

If in the subwindow for Select device a sample changer tower is selected then the

subwindow Functions/Parameters opens and shows the functions that are possi-
ble with the sample changer together with their associated parameters.

Sample changer functions

The sample changer functions are shown on the following tabs:

• General
• Move
• Assign position
• Pump
• Heater/Gas

Devices

The sample changer functions can be carried out with the following instruments:

730, 774, 778, 789, 814, 815, 855