Commands

489

Measuring parameters

The measuring parameters are defined on this tab.

Command name

25 characters, [ MEAS Cond # ]

Name of the command.

Measurement

Measurement frequency

[ auto ], 300 Hz, 2.4 kHz

The more suitable frequency is chosen with the setting

auto

.

Measuring time

0 ... [ 120 ] ... 999'999 s

Maximum period of time for measurement.

Measuring interval

1 ... [ 2 ] ... 999'999.0 s

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

0.0000 ... 2000.0 mS/cm, [ off ]

Stop when the preset measured value has been reached since the start of
the measurement.

off

means no stop.

Temperature

Reference temperature

-170.0 ... [ 20.0 ] ... 500.0 °C

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).

Measurement temperature

-170.0 ... [ 20.0 ] ... 500.0 °C

Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.