Subwindow Devices

703

Shift position

[ 0 ] ... 100 mm

Shift position for Lift 1. Each time that the rack rotates the lift will move to

this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here then the rotation will take place at
the current lift position. This means that the shift position must be selected
so that a safe rotation over the whole rack is possible at any time.

Special position

[ 0 ] ... 100 mm

Special position for Lift 1. This additional definable position can be used
e.g. during pipetting so that the tip is just immersed in the sample solution.

Special beakers

Special beaker

Shows the number of the special beaker for the selected rack.

Rack position

Shows the number of the rack position for the special beaker.

Dosing device

Shows the number and type of dosing devices connected to the 730 Sample

Changer or 774 Oven Sample Processor. A maximum of 12 type

685 Dosimat

or

700 Dosino

dosing devices can be connected.

Note

Connecting up the dosing devices is described in the Instructions for Use for the
730 Sample Changer and 774 Oven Sample Processor. If no dosing device is
connected then the card will remain empty.