Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 458

Advertising
background image

432

Method

The measured values are not entered in the list of measuring points until
the dosing rate has fallen below this value. This parameter is checked 10 s

after the start at the earliest.

Stop conditions

Conditions for stopping the titration. The stop conditions are checked at the same
time and the condition which is fulfilled first stops the titration.

Command name

25 characters, [ STAT pH # ]

Name of the command.

Stop volume

0.00000 ... [ 100.000 ] ... 9999.99 mL, off

(Titrando)

0.00 ... [ 100.00 ] ... 9999.99 mL, off

(Titrino)

Stops when the given volume (including start conditions) has been added
after the start of the titration. The stop volume should be adapted to suit the

sample weight or the titration vessel size.

Stop time

0 ... 999999 s, [ off ]

(only Titrando)

Stops when the preset time has elapsed after the start conditions have
been fulfilled.

since

Specifies the beginning of the stop time (will ot be displayed if the

Stop time

is disabled):

[ Start ]

The stop time starts after the start conditions have been fulfilled.

Control point first reached

The stop time starts as soon as the control point has been reached for the

first time.

Last dosing

The stop time starts after the last increment has been dosed, i. e. the stop
time is reset to zero again after each dosing step.

Stop rate

0.01 ... 166.00 mL/min, [ off ]

(Titrando)

0.01 ... 150.00 mL/min, [ off ]

(Titrino)

Stops when the current titration rate is lower than the rate defined here. This
parameter is checked 10 s after the start at the earliest. The maximum stop
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Filling rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Advertising