Commands

355

off

The temperature will not be measured. The

Temperature

(see Titration pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only Titrando)

If this option is enabled, the stirrer will be switched off automatically when
the command is finished.

Switch on/off automatically

[ on ], off

(only 751, 785, 798, 799)

If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command.

Start conditions

Screenshot and parameters: see DET U - Start conditions

Titration parameters

Screenshot and parameters: see DET pH - Titration parameters

Stop conditions

Screenshot and parameters: see DET U - Stop conditions

Potentiometric evaluation

Parameters for the potentiometric evaluation of the titration curve with automatic
equivalence point recognition. One of the following evaluation methods can be
chosen: