Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 438

Advertising
background image

412

Method

Note

You can use the

Delay time

to calculate the maximum

Stop drift

to be used in the

following way: the size of the last added increment depends on the volume of
the attached exchange unit. With a 20 mL exchange unit (10'000 pulses per cy-
linder volume; for KFT command is independent of the dosing drive) the smal-

lest possible increment is 2 µL. With a Delay time of 5 s the last 2 µL of reagent
must therefore suffice for 5 s or longer. This gives a drift of ≤ 2 µL/5 s ≤ 24
µL/min. (The drift may be smaller than 24 µL/min as we do not know whether the
last increment would also have been sufficient for 10 s.) This means that if you
use a 20 mL exchange unit (10'000 pulses per cylinder volume; independent of

the dosing drive) and 5 s switch-off delay then you should set a value of ≤ 24
µL/min as the stop drift.

Stop time

1 ... 999999 s

(only 720, 736, 751, 758, 784, 795, 799)

The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for

Stop criterion

=

time

and

Delay time

=

inf.

.

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ KFT Ipol # ]

Name of the command.

Titration direction

Selection of the titration direction. If two endpoints are set the direction is al-
ready given and the following selection is ignored.

+

positive measured value alteration

[ - ]

negative measured value alteration

auto

The titration direction is determined automatically from the start measured
value and the set endpoint.

Extraction time

[ 0 ] ... 999999 s

The titration is not stopped until the extraction time has elapsed (even when
the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.

Advertising