Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 409

Advertising
background image

Commands

383

Parameters

The parameters for the command

SET pH

are configured in the following 8 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

Start conditions

Parameters for the measurement of the initial measured value and for defin-
ing the conditions which have to be met at the start of the titration.

Control parameters

Control parameter for endpoint 1 and endpoint 2 (measured value for the
endpoint, titration rate, user defined parameters, stop criterion).

Titration parameters

Parameters for the run of the titration.

Stop conditions

Definitions of conditions which cause the stop of the titration.

Conditioning

Switching conditioning on/off. Definition of start drift, drift correction and

stop conditions.

Additional evaluations

Definition of additional methods for evaluation of the titrations curves.

Additional measured values

Definition of additional measured values of other measuring commands
which are saved as additional column in the list of measuring points.

General/Hardware

Screenshot and parameters: see DET pH - General/Hardware

Start conditions

The start conditions are processed in the listed sequence before the titration is
started.

Command name

25 characters, [ SET pH # ]

Name of the command.

Advertising