Cal loop ph – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 531

Advertising
background image

Commands

505

Move

:

Special beaker

Number

:

= ### + 'Command name.LCO'

(entry with Formula editor, i.e.

= 5 +

'Calibration loop pH.LCO'

for the above sample method with the first calibra-

tion solution on special beaker position

6

).

Defining rack positions for calibration solutions in the sample data

A

Sample position

for the first calibration solution has to be defined in the run

window. The other calibration solutions must be placed on the directly fol-
lowing rack positions. The following settings have to be done in the com-
mand

MOVE

under the parameter

Target

:

Move

:

Rack position

Number

:

= 'MV.Sample position' – 1 + 'Command name.LCO'

(entry with For-

mula editor, i.e.

= 'MV.Sample position' – 1 + 'Calibration loop pH.LCO'

for the

above sample method).

The variable

'Command name.LCO'

(index number of the calibration loop) is in-

creased by

+1

after each pass of the loop.

Note
The Formula editor is opened with a right click in the input field.


CAL LOOP pH

Loop command for calibrating pH electrodes. The command is used together
with the command

CAL MEAS pH

.

Appearance

The command has the following appearance:

Parameters

The parameters for the command

CAL LOOP pH

are configured in the following dia-

log window:

CAL LOOP pH