Determination parameters – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 137

Advertising
background image

Subwindow Run

111

COND READY

Conditioning criterion has been fulfilled.

COND HOLD

Conditioning is interrupted.

ERROR

Error.

Determination parameters

User

50 characters

If the option

Enforce login with user name

is switched on in the Security set-

tings then the short name of the user currently logged in will be shown in
this field. The field cannot then be edited. If work is carried out without login
then a user name can be entered in this field (only in the status

READY

).

Sample number

[ 0 ] ... 99999

At each program start the sample number will be set to

0

for all workplaces.

Each time that a determination is started it will be increased by

+1

. It can

only be edited by the user in the status

READY

.

Remark

100 characters

Freely definable remark about the determination that is saved together with

the determination. Instead of a manual entry it is also possible to select a
prepared Text template.

Note

The field

Remark

can also be edited live while determinations are being carried

out. With the context-sensitive menu item

Modify remark

the

Live modification

window opens.

Statistics

[ on ], off

This field is only shown when a method has been loaded in whose

START

command the option

Statistics

is switched on. The statistics defined in the

method can be switched on and off manually here (only in the status

READY

).

The first field after

Statistics

shows the number of determinations that have

already been carried out for the statistics (actual counter). This field cannot
be edited, but in the status

READY

the context-sensitive menu item

Delete

statistics

can be used to reset it to

0

, which at the same time deletes the

statistics data that has already been calculated.
The second field after

Statistics

shows the number of determinations to be

carried out for the statistics (set counter). It contains the value for the

Num-

ber of single determinations

defined in the

START

command as standard.

This field can only be edited in the status

READY

.

Modify remark

With the context-sensitive menu item

Modify remarks

the dialog window

Live modifi-

cations

opens, in which the remark can be edited live during a running determina-

tion.