Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 656

630
Configuration
*
= any line condition, and
p
= set pulse
The output lines and bits are numbered from right to left:
Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit
13 12 11 10 9 8 7 6 5 4 3 2 1 0
Example:
************1*
sets output line 1 to active (= set) which, e.g.,
would be a stop command for a connected Titrino.
************0*
sets
the line to inactive.
Note
We recommend that all the non-relevant output lines are masked with an asterisk
*
so that the conditions of these lines are not altered.
Pulse length
100 ... [ 200 ] ... 1000 ms
Duration of the transmitted pulse.
Examples for 765 Dosimat
Parameter
Bit-pattern
Function
Start Dos1
*******p******
Starts Dosimat 1 (Titrino via "activate")
Start Dos2
*****p********
Starts Dosimat 2 ( " )
Start Dos*
*****p*p******
Starts Dosimats 1 and 2 ( " )