Method

Sample addition time

[ 0 ] ... 99 s

Waiting time between stop of conditioning and start of titration during which

the sample can be added.

Only start titration by a start command from a SEND command

on, [ off ]

If this option is set the titration is only started when the

Event message

Start

titration

from a

SEND

command reaches this command.

Note

At the time of reception of the

SEND

command conditioning must already be

terminated (=

Condok

) otherwise the command to start the titration will be igno-

red.

Show measured value during conditioning

on, [ off ]

If this option is set additionally to the drift and time the measured value is
shown in the Live display.

Additional evaluations

The following additional methods for evaluation of titration curves can be activated
and defined here:

Command name

25 characters, [ SET pH # ]

Name of the command.

Fix end point evaluation

on, [ off ]

For a fixed quantity (measured value, volume or time) for the fix end point

the associated values are interpolated from the list of measuring points.

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other
quantities are to be interpolated from the list of measured points.

Fix EP# at pH

-20.000 ... 20.000, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.