Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 476

Advertising
background image

450

Method

Evaluations

The following methods for evaluation of titration curves can be activated and de-

fined here:

Command name

25 characters, [ STAT U # ]

Name of the command.

Rate evaluation

on, [ off ]

With this option a maximum of 9 time windows can be defined in which the

mean dosing rate can be determined by linear regression. The defined time
windows are shown in the window table and can be edited with the follow-
ing buttons:

Open the dialog window

Evaluation window #

(see below) to enter the pa-

rameters of a new time window.

Open the dialog window

Evaluation window #

(see below) to edit the pa-

rameters of the selected time window.

Delete the selected time window.