Commands

579

Command name

25 characters, [ Data request # ]

Name of the command.

Sample data request

The sample data to be entered only while the determination is running can
be selected here. The dialog window

Sample data request

with the sample

parameters defined here will appear during the method run. The sequence
is halted until the input is confirmed.

Note

In case the

sequence should not be stopped (for example for back-weighing

in KF titrations), the REQUEST command has to be inserted in another track than
the one with the titration command. The option

Return immediately

must be acti-

vated in this track.

Sample position

on, [ off ]

Switching on/off the request for the sample position.

ID1 ... ID8

on, [ off ]

Switching on/off the request for the sample identifications

ID1 ... ID8

. If

other names were given for this variables in the

START

command, these are

displayed in the dialog window

Sample data request

.

Note

The sample identifications

ID1 ... ID8

are by default of the type

Text

. If you wish

to enter numbers for this method variables for using them in later calculations,
you have to change the type to

Number

in the

START

command.

Sample size

on, [ off ]

Switching on/off the request for the

Sample size

.

Unit

on, [ off ]

Switching on/off the request for the

Sample size unit

.

Message

on, [ off ]; Text (unlimited)

This message is displayed in the dialog window

Sample data request

. The

Text editor to enter or change the message is opened with

or by

double-clicking into the message field.

Note

In case the option

Comments on modification of sample data (live)

is activated in

the Security settings, the dialog window

Modification comment sample data

(see

Subwindow Run - Single determination - Modification comment sample data) is
displayed after entering the sample data. A

Reason

and a

Comment

must be en-

tered for the modification.