Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 141

Advertising
background image

Subwindow Run

115

signed in a command then the window

Device assignment

appears for as-

signing the device.

Check solutions

Checks whether the solutions demanded in the device-specific commands
are present and available.

Check sensors

Checks whether the sensors demanded in the device-specific commands

are present and available.

Note
The Start test can also be triggered manually in the

READY

status with the menu

item

Tools, Run test

.

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

,

Statis-

tics

).
Conditioning requirement not fulfilled

After the start of preconditioning the status changes to

COND BUSY

. In this

status conditioning can be started with

[Start]

even though the conditioning

requirement has not been fulfilled. Conditioning can also be canceled with

[Stop]

or interrupted with

[Hold]

and then continued with

[Cont]

. If an error

occurs during conditioning then neither the Error track nor the Exit track will
be carried out.

Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled then the status changes to

COND READY

. In this status the determination can be started with

[Start]

.

Conditioning can also be canceled with

[Stop]

or interrupted with

[Hold]

and

then continued with

[Cont]

. The conditioning requirement will continue to be

monitored continuously until the determination has been started.

5 - Start method

At the start of the method the following checks and actions are triggered:

Check sample data

Checks that the entered Sample data corresponds to that defined in the
Method variables and is valid.

Assign determination ID

An unambiguous determination ID is assigned for the determination.

Reserve device (module)

The devices (or their functional units) used by the device-dependent com-
mands contained in the method are reserved for the duration of the deter-

mination, i.e. they cannot be used in other workplaces nor can they be op-
erated manually.

Increment sample number

The

Sample number

is increased by

+1

.

Increment statistics counter

If statistics has been switched on in both the

START

command and on the

tab

Single determinations

then the statistics counter is increased by

+1

. If a

new method is loaded or if the statistics counter agrees with the set counter
then the current statistical data will first be deleted and the statistics counter
reset to

0

.