Commands

457

Sensor

Measuring input

[ 1 ] ... 2

(Titrando)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ pH electrode ]

Selection of a sensor of type

pH electrode

from the list of sensors defined in

the sensor table.

Temperature measurement

Type of temperature measurement (Titrando only):

continuous

A temperature sensor must be connected. The temperature is then meas-
ured continuously.

[ automatic ]

If a temperature sensor is connected, the temperature will be measured

continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off

The temperature will not be measured. The temperature manually entered
under

Temperature

(see Measuring parameters) will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(Titrando only)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(Titrando only)

Setting the stirrer speed. The direction in which stirring takes place changes

as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(Titrando only)

The stirrer is automatically switched off at the end of the command if this
option is enabled.