552

Method

Gas

Gas type

[ air ], nitrogen, other gas

Gas type of the carrier gas.

Gas flow factor

0.001 ... [ 1.000 ] ... 9.999

Factor for the correct determination of the gas flow. This parameter is only
visible for

Gas type

=

other gas

. The values for frequently used carrier gases

are listed in the Instructions for Use for the 774 Oven Sample Processor.

Gas flow

Inlet

Selection of the inlet of the carrier gas:

[ pump ]

Use of ambient air as carrier gas.

valve

Use of gas out of a compressed gas cylinder as carrier gas.

Switch on

[ on ], off

Switch on the gas flow.

Switch off

on, [ off ]

Switch off the gas flow.


5.6.10 Result

commands

Commands for the calculation, storage and output of results.
The following four result commands can be selected:

CALC

Calculation of intermediate and end results, titer values and common vari-
ables.

DATABASE

Storage of the determination data in databases.

EXPORT

Export of determination data.

REPORT

Output of a report defined by a report template.

CALC

Command for the calculation of intermediate and end results and the as-
signment of titer values and common variables.

Appearance

The command has the following appearance: