Commands
437
Track name
Selection of available tracks
Selects the track that is to be started automatically.
Note
If a track is called for that is already running then it will be allowed to finish and
then restarted when it is free again.
Evaluations
The following methods for evaluation of titration curves can be activated and de-
fined here:
Command name
25 characters, [ STAT pH # ]
Name of the command.
Rate evaluation
on, [ off ]
With this option a maximum of 9 time windows can be defined in which the
mean dosing rate can be determined by linear regression. The defined time
windows are shown in the window table and can be edited with the follow-
ing buttons:
Open the dialog window
Evaluation window #
(see below) to enter the pa-
rameters of a new time window.