764

Configuration

Beaker radius

Radius of the special beaker.

Beaker sensor

Shows whether and which beaker sensor is to be used for the special
beaker.

Opens the dialog window

Special beaker #

for editing the data of the se-

lected special beaker.

Special beaker

In the dialog window

Special beaker #

the parameters for the special beaker se-

lected in the Table of special beakers can be defined.

Rack position

[ 0 ] ... n (n is rack-dependent)

Number of rack position for selected special beaker.

Note

Special beakers should preferably be set at high rack positions, so that a sam-
ple series can start at rack position 1. Rack positions that are defined as special
beakers are no longer available as sample positions. They will be jumped over
during an automatic movement to a sample position.

Work position Tower 1

[ 0 ] ... 235 mm

Work position of the selected special beaker at Tower 1.

Work position Tower 2

[ 0 ] ... 235 mm

Work position of the selected special beaker at Tower 2.

Beaker radius

1.0 ... 100.0 mm, [ off ]

Radius of the selected special beaker on the rack. If the lift is moved to the
work position then this value will be compared with the parameter

Min.

beaker radius

(see Properties Tower #) that can be specifically defined for

each tower. If Beaker radius samples < Min. beaker radius then a cor-
responding error message will appear. With

off

the beaker radius will not be

checked.

Beaker sensor

[ Tower ], Robotic arm, off

When the selected special beaker is moved to with the command

MOVE

then the beaker sensor (

Tower

,

Robotic arm

) checks whether the special

beaker is present or not. With

off

no check will be made. For the option

Ro-

botic arm

a swing head with beaker sensor must be installed and a suitable

work position with beaker contact must be defined for the lift, as this must

move to the beaker recognition position.
With the parameter

Beaker test

in the command

MOVE

you can determine

whether the determination is to be terminated if a beaker is missing with or
without the display of a corresponding message and whether the series is
to be continued or also terminated.