Configuration data

629

Note

Input lines that are not of interest, or for which no defined condition can be pre-
dicted, should be masked with an asterisk

*

.

Templates for output lines

With the menu item

Tools, Templates, Output lines

the dialog window

Templates for

output lines

opens in which the client-specific bit-pattern for setting remote output

signals can be defined; these can be selected with the command

CTRL

and in

Manual Control.

Template table

The table with the defined templates cannot be edited, however, with a mouse
click on the column title it can be sorted according to the selected column in ei-
ther an increasing or decreasing sequence.

Generates a new template (see below).

Edits the selected template (see below).

Deletes the selected template.

Generate new template or edit template

With

<New>

or

<Edit>

the dialog window

New template

or

Edit template

opens for

entering a new template or for editing an existing template respectively.

Signal name

25 characters

Name of template for the output signal.

Output signal

Bit-pattern with exactly 14 characters (0, 1, *, p), [ ************** ]

Entry of the bit-pattern for the output signal with exactly 14 characters.
The following characters can be entered

0

= line inactive,

1

= line active,