Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 367

Advertising
background image

Commands

341

Calculated measured value

With the option

Additional calculated measured values

3 new measured values can

be calculated from existing measured values. They can be displayed in curves
with the identification

Calculated 1...3

or used as variables

'Command name.CA1...3'

in formulae.

Name

25 characters, [ Calc. measured value # ]

Designation for the calculated measured value. This name is used as axis
legend.

Formula

1000 characters, [ 'empty' ]

Formula to calculate the measured value. It can either be entered directly or

can be created in the formula editor which opens with

.

Unit

16 characters, [ 'empty' ]

Unit of calculated measured value.

Decimal places

0 ... [ 2 ] ... 5

Number of decimal places with which the calculated measured value will be
displayed.

Comment

250 characters, [ 'empty' ]

Comment about the calculated measured value.

Additional external measured values

on, [ off ]

3 new measured values can be taken over from existing measuring com-
mands (i.e.

MEAS

) running simultaneously in other tracks. They can be dis-

played in curves with the identification

External 1...3

or used as variables

'Command name.EX1...3'

in formulae.

Open the dialog window

External measured value #

(see below) with the pa-

rameters for the external measured value.

Open the dialog window

External measured value #

(see below) to edit the

parameters of the external measured value.

Delete the selected external measured value.

Advertising