Templates for input lines – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 654

Advertising
background image

628

Configuration

Templates for input lines

With the menu item

Tools, Templates, Input lines

the dialog window

Templates for

input lines

opens in which the client-specific bit-pattern for scanning remote input

signals can be defined, these can be selected with the command

SCAN

and in

Manual Control.

Template table

The table with the defined templates cannot be edited, however, with a mouse
click on the column title it can be sorted according to the selected column in ei-
ther an increasing or decreasing sequence.

Generates a new template (see below).

Edits the selected template (see below).

Deletes the selected template.

Generate new template or edit template

With

[New]

or

[Edit]

the dialog window

New template

or

Edit template

opens for en-

tering a new template or for editing an existing template respectively.

Signal name

25 characters

Name of the template for the input signal.

Input signal

Bit-pattern with 8 characters (0, 1, *), [ ******** ]

Entry of the bit-pattern for the input signal with exactly 8 characters.

The following characters can be entered

0

= line inactive,

1

= line active, and

*

= any line condition.

The input lines and bits are numbered from right to left:

Input 7 6 5 4 3 2 1 0

Bit

7 6 5 4 3 2 1 0

Example:

*******1

expects an active input line 0 (

1

= set). This line will be

set, e.g. by a Titrino when a titration has been finished and the Titrino can

again receive a start signal.

Advertising