724

Configuration

Action

If during solution monitoring it is found that the working life has expired then one

of the following actions will be triggered automatically at Start test:

Record message

The message that the working life of the solution has expired will be auto-
matically saved with the determination.

Display message

A message will appear and you can choose whether to continue the run or
terminate it. If the run is continued then the message that the working life of

the solution has expired will be automatically saved with the determination.

Stop determination

The running determination will be automatically terminated. The following
message must be confirmed with

[OK]

.

Titer

Titer determination

Titer

Value:

1E-12 ...

[ 1.000 ] ... 1E+12, max. 10 characters

Unit:

10 characters, [ 'empty' ], mol/L, mmol/L, µmol/mL, g/L, mg/L, µg/mL,

mg/mL, ppm, %, mEq/L

Titer of the solution. The value for the titer of a solution used by the titration
commands

DET

,

MET

,

SET

and

KFT

is available as a variable

'Command

name.TITER'

for calculations.

Date titer det.

Date and time of last titer determination, which is automatically entered af-

ter each automatic titer determination or manual entry.

Titer method

Name of the method with which the last titer determination was carried out.
If the titer has been entered manually then

manual

will be shown here.

User

Short name of the user logged in during titer determination or when the titer

was entered manually. If work is not carried out using log in then the user
logged in under Windows will be entered automatically.