Commands

387

Note

You can use the switch-off time in the following way to calculate the maximum
stop drift to be used: the size of the last increment to be added depends on the
volume of the attached exchange unit. With a 20 mL exchange unit (10'000 pul-
ses per cylinder volume) – number of impulses depends on the dosing device –

the smallest possible increment is 2 µL. With a delay time of 5 s this means that
the last 2 µL of reagent to be added must be sufficient for 5 s or longer. This re-
sults in a drift of ≤ 2 µL/5 s ≤ 24 µL/min. The drift can be smaller than 24 µL/min
as we do not know whether the last increment would also have been sufficient
for 10 s. This means that if you have previously worked with a 20 mL exchange

unit (10'000 pulses per cylinder volume) and 5 s switch-off delay then you should
set a value of ≤ 24 µL/min as the stop drift.

Stop time

1 ... 999999 s

(only Titrino)

The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for

Stop criterion

=

time

and

Delay time

=

inf.

Endpoint 2

(only visible for

EP1 at pH

off

)

[ on ], off

Switching endpoint 2 on/off. If this is deactivated all following parameters
are not visible.

EP2 at pH

-20.000 ... 20.000, [ off ]

(Titrando),

-20.00 ... 20.00, [ off ]

(Titrino)

Measured value for the second endpoint.
Parameters for endpoint 2 see above.

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ SET pH # ]

Name of the command.

Titration direction

Selection of the titration direction. If two endpoints are set the direction is al-

ready given and the following selection is ignored.

+

positive measured value alteration

[ - ]

negative measured value alteration

auto

The titration direction is determined automatically from the start measured

value and the set endpoint.