Metrohm tiamo 1.2 Manual Benutzerhandbuch
Seite 427

Commands
401
Start volume
Start volume
[ 0.00000 ] ... 9999.99 mL
(Titrando),
[ 0.00 ] ... 999.99 mL
(Titrino)
Volume to be added before the start of the titration with the below entered
speed.
Dosing rate
0.01 ... 166.00 mL/min, [ maximum ]
(Titrando),
0.01 ... 150.00 mL/min, [ maximum ]
(Titrino)
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.
Pause 2
Pause 2
[ 0 ] ... 999999 s
Waiting time, e.g. until the electrode has settled down after the start or a re-
action time after the addition of a start volume. The pause takes place at
the end of all start conditions.
Control parameters
The Control parameters can be set for each endpoint separately.
Command name
25 characters, [ SET Upol # ]
Name of the command.
Endpoint 1
EP1 at
-200.0 ... 200.0 µA, [ off ]
Measured value for the first endpoint.
Titration rate
Titration rate
Three predefined sets of parameters can be selected for the titration rate:
slow
,
optimal
and
fast
. For these sets the parameters for the
Control
are not