Commands

357

Fixed quantity

[ Measured value ], Time, Volume

Selection of the fixed quantity to which the associated values for the other

quantities are to be interpolated from the list of measured points.

Fix EP# at

-2000.0 ... 2000.0 mV, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Measured value

.

Fix EP# at

0.0 ... 999999.9 s, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Time

.

Fix EP# at

0.00000 ... 9999.99 mL, [ off ]

Fix end point 1 ... 9 for

Fixed quantity

=

Volume

.

With this button additionally to the always shown fix EP1...3 the fix EP4...9
are displayed.

With this button instead of fix EP1...9 only the fix EP1...3 are shown.

pK/HNP evaluation

on, [ off ]

With this evaluation from the titration curve the pK value can be determined
.which corresponds to the pH value at the half neutralization point (see Ti-
tration commands - Evaluation - pK value and half neutralization potential
).

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 25.0 ] ... 2000.0 mV/mL

The evaluation begins as soon as the slope of the curve exceeds the set

threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.1 ... [ 25.0 ] ... 2000.0 mV/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Break point evaluation

on, [ off ]

(only Titrando)

With the break point evaluation sharp changes of direction in the titration

curve are determined (see Titration commands - Evaluation - Breakpoint
evaluation
).

EP criterion

0 ... [ 0.3 ] ... 1.0

Measure of the minimum sharpness of the break point. The lower the EP
criterion set, the more break points will be found. As this is a relative value

related to the total measured value alteration, for a small measured value

Advertising