Subwindow Rack data

761

Rack parameters

Beaker radius samples

1.0 ... 100.0 mm, [ off ]

Radius of the sample beakers on the rack. If the lift is moved to the work
position then this value will be compared with the parameter

Min. beaker ra-

dius

(see Properties Tower #) that can be specifically defined for each

tower. If

Beaker radius samples < Min. beaker radius

then a corresponding er-

ror message will appear. With

off

the beaker radius will not be checked.

Beaker sensor

[ Tower ], Robotic arm, off

When a sample position is moved to with the command

MOVE

then the

beaker sensor (

Tower

,

Robotic arm

) checks whether the special beaker is

present or not. With

off

no check will be made. For the option

Robotic arm

a

swing head with beaker sensor must be installed and a suitable work posi-
tion with beaker contact must be defined for the lift, as this must move to
the beaker recognition position.
With the parameter

Beaker test

in the command

MOVE

you can determine

whether the determination is to be terminated if a beaker is missing with or

without the display of a corresponding message and whether the series is
to be continued or also terminated.

Rack offset

-5.00 ... [ 0.00 ] ... 5.00 °

The rack offset is a production-dependent tolerance value between the up-
per part of the rack and the lower part. The rack offset value is determined

during rack adjustment and shown here. It can be edited if necessary.