Commands

495

Addition mode

Fixed volume increments

[ on ], off

If this option is enabled a fixed volume increment muss be added for each

standard addition.

Addition volume

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the fixed volume increment.

Variable volume increments

on, [ off ]

If this option is enabled the defined volume increment muss be added for

each standard addition.

Addition volume 1...19

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the defined volume increments.

Addition solution

Concentration

0.001 ...[ 1.000 ] ... 999999.999

Concentration of addition solution.

Unit

mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L

Concentration unit of addition solution.

Measuring parameters

The measuring parameters are defined on this tab.

Command name

25 characters, [ STDADD man # ]

Name of the command.

Measurement with drift control

The measurement is stopped as soon as the

Signal drift

or the

Stop meas-

ured value

is reached or a

waiting time

has elapsed.

Signal drift

0.1 ... [ 10.0 ] ... 999.0 mV/min

The measured value is only accepted if the signal drift defined here has
been fallen short of.