566

Method

CTRL - Properties

Command name

25 characters, [ Set lines # ]

Name of the command.

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able
to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Remote box

[ 1 ] ... 4

(only Titrando, 814, 815)

Selection of remote box at which the lines are to be set.

Set lines

Output signal

Binary pattern of exactly 14 characters (0, 1,*, p), [ ************** ], signal
template

Input of the binary pattern for the output signal or selecting a predefined

Signal pattern.
The following characters can be entered:

0

= Line inactive

1

= Line active

*

= Any line condition

p

= Set impulse (impulse length 200 ms. If an impulse with another length

should be emitted, a corresponding template must be defined.)
The bits are always numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example:

************1*

sets output line 1 to active (= set), for example,

with a connected Titrino this would be a Stop command.

************0*

resets the output line to inactive.