Method

Start conditions

The start conditions are processed in the listed sequence before the titra-

tion is started.

Command name

25 characters, [ SET Upol # ]

Name of the command.

Initial measured value

The initial measured value is determined before the other start conditions are
processed. The measured value acceptance depends on the three following pa-
rameters:

Signal drift

0.1 ... 99.90 µA/min, [ off ]

(only Titrando)

The measured value is only accepted if the drift undercuts the entered
value.

off

means that the measured value is accepted after the maximum

waiting time.

Min. waiting time

[ 0 ] ... 999999 s

(only Titrando)

The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is accepted only after the minimum waiting
time has elapsed, even if the signal drift has already been achieved. The
drift continues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 1 ] ... 999999 s

(only Titrando)

If signal drift has been switched off or has not yet been achieved then the
measured value will be accepted when the maximum waiting time has
elapsed

Pause 1

Pause 1

[ 0 ] ... 999999 s

(only Titrando and Titrino 719, 720, 736, 751, 758, 785, 794,

798, 799)
Waiting time, e.g. until the electrode has settled down, before a start vol-
ume is added.