424

Method

Control range

Set a large control range for steep curves and a small control range for flat ones.

A good approximation for the start of the control range is given by the point where
the tangents intersect.


STAT pH

Command for STAT titrations with measured value pH.

Devices

This command can be executed with the following devices which have the

STAT

pH

mode:

Titrino: 718, 736, 751, 799

Titrando: 835, 836, 842, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STAT pH

are configured in the following 8 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.

Start conditions

Parameters for the measurement of the initial measured value and for defin-
ing the conditions which have to be met at the start of the titration.

Control parameters

Parameters for the control point.

Titration parameters

Parameters for the run of the titration.

Stop conditions

Definitions of conditions which cause the stop of the titration.

Monitoring

Definition of measured value, dosing rate, and temperature monitoring.