Administration

601

Login/Password protection

On the tab

Login/Password protection

is defined whether the user must log in with

his/her name or with name and password, and how the password is to be moni-
tored and constructed.

Settings according to 21 CFR 11

Sets the parameters to default settings for FDA-conform work.

Login policies

Enforce login with user name

on, [ off ];

FDA-default:

on

If this option is switched on then every time that the program starts the

login window will appear, in which the user must enter his/her user name. If
this option is switched off then no login will be demanded and the user
name logged in under Windows will be adopted. In this case all subsequent
parameters are inactive.

Enforce login with password

on, [ off ];

FDA-default:

on

If this option is switched on then every time that the program starts the
login window will appear, in which the user must enter his/her user name
together with a password. If this option is switched off then all subsequent
parameters are inactive.

Password monitoring by tiamo

on, [ off ];

FDA-default:

on

If this option is switched on then the password will be monitored by tiamo
according to the following parameters.

Password monitoring by Windows

on, [ off ];

FDA-default:

off

If this option is switched on then the password will be monitored accord-
ing to the parameters defined in Windows. In this case the parameters for
password protection are inactive.