Subwindow Run

125

Check device (module) and rack

Checks whether the devices (or device types), modules (e.g. measuring in-

puts, towers, etc.) and racks demanded in the device-specific commands
are present and available or assigned. If the device has not yet been as-
signed in a command then the window

Device assignment

appears for as-

signing the device.

Check solutions

Checks whether the solutions demanded in the device-specific commands
are present and available.

Check sensors

Checks whether the sensors demanded in the device-specific commands
are present and available.

Check sample data

Checks whether the entered sample data corresponds to that defined in
method variables and is valid.

Note

The Start test for the number of lines defined under

Autostart

can also be trigge-

red manually in the

READY

status with the menu items

Tools, Run test

or

Sample

table, Run test

.

4 - Preconditioning (only for methods with conditioning)

In methods that contain commands with switched-on conditioning and in which
the option

Automatic conditioning

is switched on in the

START

command then pre-

conditioning will start for all these commands. Preconditioning does not belong to
the determination itself and does not increase any counter (

Sample number

or

Sta-

tistics

).

Conditioning requirement not fulfilled

After the start of preconditioning the status changes to

COND BUSY

. In this

status conditioning can be interrupted with

[Hold]

and continued with

[Cont]

. If an error occurs during conditioning then neither the Error track nor

the Exit track will be carried out.

Conditioning requirement fulfilled

If the conditioning requirement has been fulfilled the status changes to

COND READY

and the method then starts automatically.

5 - Start method

At the start of the method the following checks and actions are triggered:

Assign determination ID

An unambiguous determination ID is assigned for the determination.

Reserve device (module)

The devices (or their functional units) used by the device-dependent com-
mands contained in the method are reserved for the duration of the deter-
mination, i.e. they cannot be used in other workplaces nor can they be op-

erated manually.

Increment sample number

The

Sample number

is increased by

+1

.

Increment statistics counter

If statistics has been switched on in both the

START

command and on the

tab

Determination series

then the statistics counter will be increased by

+1

. If

a new method is loaded or if the statistics counter agrees with the set
counter then the current statistical data will first be deleted and the statistics
counter reset to

0

.