Manual control

65

Add fixed volume

Adding a predefined volume can be started and stopped here. This tab only ap-
pears when a single dosing device has been selected.

Volume

[ 0.100 ] ... 99999.9 mL

Fixed volume to be added.

Dosing rate

[ maximum ]
0.01 ... 166.00 mL/min

(Titrando, 814, 815, 855)

0.01 ... 160.00 mL/min

(730, 774, 778, 789)

0.01 ... 150.00 mL/min

(Titrino)

The volume is added at this speed. The maximum dosing speed depends
on the cylinder volume of the Exchange unit or Dosing unit used. If the en-
tered dosing rate is too high for the selected dosing device then during
dosing it will automatically be reduced to the largest possible value.

Note
The dosing rate should be reduced for viscous liquids.

Filling rate

[ maximum ]
0.01 ... 166.00 mL/min

(Titrando, 814, 815, 855)

0.01 ... 160.00 mL/min

(730, 774, 778, 789)

0.01 ... 150.00 mL/min

(Titrino)

After dosing the buret will be refilled at this speed. The maximum filling
speed depends on the cylinder volume of the Exchange unit or Dosing unit
used. If the entered filling rate is too high for the selected dosing device
then during filling it will automatically be reduced to the largest possible

value.