9 automation commands, Move – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 564

Advertising
background image

538

Method

Dosing device

Dosing device

[ 1 ] ... 4

(Titrando, 846)

[ 1 ] ... 3

(814, 815, 855)

Selection of the dosing device (Dosinos only) to be emptied. All the dosing
device connections (MSB) are displayed which are possible with the se-
lected device type.

Solution

24 characters, solution name, [ not defined ]

Input or selection of the solution defined in the solution table. Additionally,

not defined

can be selected. If intelligent exchange or dosing units are used

then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder

volume is checked. At the start of the command the validity of the titer and
the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If

not defined

is selected no tests will be carried

out.


5.6.9 Automation

commands

Commands for the operation of Sample Processors.

The following eight automation commands can be selected:

MOVE

Moving to a rack position or an external position.

SWING

Swinging of the robotic arm (with Swing Head only).

LIFT

Moving to a lift position.

PUMP

Switching on/off the connected or built-in pumps.

STIR

Controlling a connected stirrer.

RACK

Initialization of the rack attached.

HEATER

Controlling the oven temperature of the 774 Oven Sample Processor.

FLOW

Regulation of the gas flow of the 774 Oven Sample Processor.

MOVE

Command for moving to a rack position (sample position or special beaker).

Devices

This command can be executed with the following devices:

Sample Processor: 730, 774, 778, 789, 814, 815, 855