Manual control

71

2.6.7 Remote

functions

If in the subwindow for Select device a remote box or remote interface is selected
then in the subwindow for Functions/Parameters the functions and parameters
that are possible for the device will appear.

Devices

The remote functions can be carried out with the following instruments:

Titrino: 702*, 716*, 718*, 719*, 720*, 721*, 736*, 751, 758, 784, 785, 794*, 795,
798, 799 (* instruments with only 3 outputs)
Titrando: 808, 809, 835, 836, 841, 842, 846, 855

Coulometer: 756, 831

Sample Changer: 730, 774, 778, 789, 814, 815, 855

Inputs

Current status

Shows the current status of the 8 input lines.

Templates/Entry

If the current status corresponds to one of the defined Templates for the in-
puts then the corresponding name will be shown here.

Note
Only templates without stars

*

are recognized.

Outputs

Current status

Shows the current status of the 14 output lines.

Templates/Entry

Binary pattern consisting of exactly 14 characters (0, 1, *, p), [
************** ], signal template

Input of the binary pattern for the output signal of exactly 14 characters or
selects a predefined Signal template.
The following characters can be entered:

0

= line inactive,

1

= line active,

*

= any line status

p

= set impulse (not for 730, 774, 778, 789). The impulse length is 200 ms.