Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 596

Advertising
background image

570

Method

SEND - event messages

Receiver

[ System ], Command

Selection of the type of receiver.

Selection

Command name, [ 'empty' ]

Selection of the command which a message should be sent to. All the
commands defined in the method are available for

Receiver

=

Command

.

For

Receiver

=

System

this parameter is not editable.

Event message

[ Quit ], Hold, Stop

(for

Receiver

=

System

)

[ Quit ], Hold, Continue, Start Titration

(for

Receiver

=

Command

)

Selection of the event message to be sent. The following event messages
can be selected depending on the type of receiver:

Receiver

Selection

Event message

Meaning

System

-

Hold

Hold determination (all tracks). Corresponds

to the button

[Hold]

in the dialog window Run.

System

-

Quit

Cancel determination (Series continues).

System

-

Stop

Cancel determination and series. Corresponds
to the button

[Stop]

in the dialog window Run.

Command

All track com-

mands

Hold

Hold selected track.

Command

All track com-

mands

Continue

Continue selected track.

Command

All track com-

mands

Quit

Cancel selected track.

If the selected track contains a command
being conditioning, conditioning is only
stopped if the option

Automatic conditioning

is

disabled in the command

START

.

Command

All commands

Hold

Hold selected command. Also valid for condi-
tioning.

Command

All commands

Continue

Continue selected command. Also valid for
conditioning.

Command

All commands

Quit

Cancel selected command.
If the selected command is being condition-

ing, conditioning is only stopped if the option

Automatic conditioning

is disabled in the com-

mand

START

.

Command

SET, KFT, KFC

Start Titration

Start titration. For this, the option

Only start

titration by a start command from a SEND com-
mand

must be enabled on the tab

Conditioning

of the titration command (Example: KFT).