Meas cond – Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 513

Advertising
background image

Commands

487

[ automatic ]

If a temperature sensor is connected, the temperature will be measured

continuously. Otherwise the temperature manually entered under

Tempera-

ture

(see Measuring parameters) will be used.

off

The temperature will not be measured. The temperature manually entered
under

Temperature

(see Measuring parameters) will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

Setting the stirrer speed. The direction in which stirring takes place changes
as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

The stirrer is automatically switched off at the end of the command if this

option is activated.

Measuring parameters

Screenshot and parameters: see MEAS U - Measuring parameters

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values

MEAS Cond

Command for conductivity measurements.

Devices

This command can only be executed with the 712 Conductometer.

Appearance

The command has the following appearance:

Parameters

The parameters for the command

MEAS Cond

are configured on the following 4

tabs:

General/Hardware

Parameters for device, sensor and stirrer.

Measuring parameters

Parameters for the measurement.

Evaluations

Defining further methods for evaluating measurement curves.

Additional measured values

Defining further measured values which can be saved as additional col-
umns in the list of measuring points.