Determination parameters

User

50 characters

If the option

Enforce login with user name

is switched on in the Security set-

tings then the short name of the user currently logged in will be shown in
this field. The field cannot then be edited. If work is carried out without log
in then a user name can be entered in this field (only in the status

READY

).

Sample number

[ 0 ] ... 99999

At each program start the sample number will be set to

0

for all workplaces.

Each time that a determination is started it will be increased by

+1

. It can

only be edited by the user in the status

READY

.

Remark

100 characters

Freely definable remark about the determination that is saved together with
the determination. Instead of a manual entry it is also possible to select a
prepared Text template.

Note

The field

Remark

can also be edited live while determinations are being carried

out. With the context-sensitive menu item

Modify remark

the

Live modifications

window opens.

Autostart

1 ... 999, [ Sample table ]

Number of automatic, internal starts of determinations within a sample se-
ries.
The first field after

Autostart

shows the number of determinations that have

already been carried out (actual counter). This field cannot be edited, but in
the status

READY

or

PAUSE

the context-sensitive menu item

Reset autostart

counter

can be used to reset it to

0

.

The second field after

Autostart

shows the total number of determinations to

be carried out per series (set counter). It contains the entry

Sample table

as

standard which indicates the number of lines of the working sample table.
This field can only be edited in the status

READY

.

Note

The second field after

Autostart

can also be altered live while determinations are

being carried out. With the context-sensitive menu item

Modify autostart counter

the window

Live modifications

opens.

Statistics

[ on ], off

This field is only shown when a method has been loaded in whose

START

command the option

Statistics

is switched on. The statistics defined in the

method can be switched on and off manually here (only in the status

READY

).