Metrohm tiamo 1.2 Manual Benutzerhandbuch

Seite 164

Advertising
background image

138

Workplace

With

or a double-click on the text field the Text editor starts with which

the message can be entered and edited. Within the text field the Formula
editor can also be called up.

Stop series ([Stop])

[ on ], off

If this option is selected then the series will be stopped before the start of
the marked line (corresponds to

[Stop]

).

Data import

Switches the import of external data on and off for working sample table fields.

Import from balance

on, [ off ]

If this option is switched on then external data from the selected balance

will be imported into the working sample table. Data can be imported into
the fields

ID1...ID4

,

Sample size

and

Sample size unit

.

Note

In order to be able to import sample identifications from the balance the data
export must be set accordingly on the balance (use names

ID1

,

ID2

,

ID3

,

ID4

).

Balance

Device name

Selection of the balance from which data is to be imported.

Always accept data

[ on ], off

If this option is selected then the data transmitted from the balance will be
imported into the working sample table for which this balance has been de-

fined as the data source.

Accept data only for active window

on, [ off ]

If this option is selected then the data transmitted from the balance will only
be imported into the working sample table of the workplace which has the
focus.

Import from barcode reader

on, [ off ]

If this option is selected then external data from the selected barcode
reader will be imported in the working sample table. The field into which the
data is to be imported is defined in the device properties of the Barcode
reader.