Command for STAT Titrations (

STAT

).

Principle

In STAT titrations reagent is added until a preset measured value has been
reached (control point); this is then kept constant. The substance released by the
reaction is immediately titrated with the reagent to maintain the control point. The

STAT titration keeps the control point constant until the preset termination criterion
has been achieved.

STAT titrations are used, for example, in enzyme analysis where the enzyme activ-
ity can be determined from the resulting dosing rate (in mL/min).

Reagent addition takes place in three phases during the titration:

Initial addition

During this phase the dosing rate increases continuously. It starts with the

Min. rate

and increases to the

Max. rate

.

Continuous addition

In this phase addition is carried out at the

Max. rate

until the

Control range

is

reached.

Control range

In this range addition is finely controlled. Shortly before the end point is
reached addition is only carried out at the

Min. rate

(see "Control range").

Commands

Depending on the measured value one of the following two

STAT

commands can

be chosen:

STAT pH

STAT Titration with pH electrodes (measuring quantity pH).

STAT U

STAT Titration with metal electrodes (measuring quantity U).