Commands

417

Generator electrode

Generator type

[ without diaphragm ], with diaphragm

Selection of the generator type.

Generator current

100, 200, [ 400 ], mA, auto

Polarization current at the generator electrode.

auto

means that the current

is automatically adapted to the conductivity of the reagent and that in the
region of the endpoint the current will be controlled at smaller values.

Note

For generator electrodes

with diaphragm

it is recommended to set the

Generator

current

to

auto

.

For generator electrodes

without diaphragm

it is recommended to set the

Gene-

rator current

to

400 mA

.

Stirrer

Switch on/off automatically

[ on ], off

If this option is eanbled, the stirrer will be switched on automatically at the

start of the command and will be switched off automatically at the end of
the command.

Start conditions

Command name

25 characters, [ KFC # ]

Name of the command.

Pause

Pause

[ 0 ] ... 999999 s

Waiting time before the start of the titration.