496

Method

Min. waiting time

[ 0 ] ... 999'999 s

The measured value is only accepted if the minimum waiting time has

elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 52 ] ... 999'999 s

If the signal drift has not yet been reached then the measured value will be
accepted when the maximum waiting time has elapsed. If no new entry for

the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

Time interval for entering a measuring point in the list of measuring points.

Temperature

Temperature

-20.0 ... [ 25.0 ] ... 150 °C

Manually entered measuring temperature. If a temperature sensor is con-

nected and

temperature measurement

on the tab

General/Hardware

is set to

automatic

or

continuous

, the temperature is measured continuously. The

value is used for the temperature adjustment during the measurement.

Switch off stirrer during measurement

on, [ off ]

If this option is switched on then the stirrer is switched off during the meas-

urement. In this case the two following parameters can be edited:

Stir before measurement

[ 0 ] ... 999999 s

Stirring time between end of dosing and measurement.

Pause before measurement

[ 0 ] ... 999999 s

Waiting time before the measurement during which no stirring takes place.

STDADD dos

Command for Standard addition with addition of the standard addition solution
from a dosing device.

Devices

This command can be executed with the following devices:

Titrando: 808, 809, 835, 836, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STDADD dos

are configured on the following 3

tabs:

General/Hardware

Parameters for device, dosing device, sensor and stirrer.