631

6.3.4 Options

Under

Tools, Options

general program properties can be set on the following 3

tabs:

General

Selection of the dialog language and switches the emergency stop button

on/off.

Save

Saves settings on exiting tiamo.

PDF

Settings for pdf files

General

Dialog language

Dialog language

Deutsch, [ English ]

Selection of the dialog language.

Note
For the altered setting to become effective tiamo must be restarted.

Emergency stop button

on, [ off ]

If this option is enabled then the emergency stop button will be shown in all
program parts.

This button can be moved to any position with the left-hand mouse key
pressed down and is shown right at the top on the screen both in the tiamo
program window and outside it on the Windows desktop. A mouse click on
this button immediately stops all running determinations on all active work-
places. The emergency stop is effective even if no user is logged in, e.g.

when the user has been logged out automatically (see Login/Password pro-
tection).

Note

The emergency stop stops all commands and tracks on all workplaces but no
device functions switched on by a command (e.g. pumps, valves, stirrers). To
stop these devices a separate track containing the commands to stop the desi-
red functions must be called in the Exit track with the condition

SV.STO = 1

. The

emergency stop has no effect on actions still running in the manual control. The-
se actions must be stopped there one by one with the

<Stop>

button.